Fourier-Transform Infrared Spectroscopy is referred to as FTIR. FTIR is a method for obtaining an infrared spectrum of an object's absorption, be it a solid, liquid, or gas. The basic idea behind FTIR is to examine a particular mixture or sample, identify chemicals, and discover functional groups. Spectroscopy is the study of interactions between chemical elements and electromagnetic radiation as well as electronic stimuli, molecular vibrations, and nuclear spin orientation. Infrared (IR) or FTIR spectroscopy is a type of spectroscopy (Fourier Transform Infrared spectroscopy).

This spectroscopy is based on a molecule's vibrations. Infrared spectroscopy is a method for simultaneously observing the interactions of molecules with various wavelengths of electromagnetic radiation. The FITR operating principles are frequently used in a variety of industries to achieve specific goals and objectives. Overall, FTIR is generally applied in the pharmacy industry, as well as in the fields of polymers & packaging, petrochemicals, forensics, oil & gas, drug abuse, research materials, and many other areas.

Fourier Transform Infrared (FT–IR) spectrometry was developed in order to overcome the limitations of dispersive instruments. A solution was developed using a very simple optical device called an interferometer. The interferometer generates a unique type of signal which has all infrared frequencies “encoded” into it. Signals can be measured very quickly. Thus, the required analysis time per sample is reduced to a few seconds. The interference wave is produced in an interferometer. The Fourier transformation is carried out, data is collected and stored, and the interferometer is controlled by a computer. The computer also does post-spectroscopic tasks like spectrum display, resolution enhancement, calibration, and the calculation of correlation equations.

The IR source projects a collimated light beam toward the Michelson interferometer, where the beam splitter divides it. From a fixed mirror, one half of the beam is reflected, and from a moving mirror, the other half.

After returning from the mirrors, the two light beams combine once more and create a beam that is optically an interference wave. Sample interaction modifies the interference light beam as it travels through the sample. The interference light beam passes through the sample and is modified by its interaction with the sample. The changed light is often detected with a Deuterated Triglycine Sulphate (DTGS) pyroelectric detector. An Analog to Digital Converter (ADC) converts the analogue signals that arrive at the detector into digital signals that are then stored in the computer. For this procedure to go well and laser technology is required. In the interferometer, a laser beam experiences the same change in optical path as an IR beam. As a result, it can be used as a guide to determine in which the mirror should be during the scan. As a result of the mirror's movement, the laser beam also controls the data collecting.

Using the quick Fourier transform approach, an entire IR spectrum can be rebuilt in just one cycle of mirror movement. Additional scans are typically conducted if time is not an issue. An average interferogram is determined following each addition. An electrical signal is represented as an interferogram as a function of the time delay (retardation) between the two beams in the interferometer. Considering that the square root of the number of scans is inversely correlated with the noise level, the addition operation increases the signal-to-noise ratio. After that, additional processing including zero filling, phase correction, and apodization are used to create the average interferogram. For the purpose of compensating for instrumental artefacts, various mathematical procedures must be used.